Concrete freeze-thaw cycle test chamber
TPDR-10/TPDR-16/TPDR-28
It is used to determine the resistance of concrete specimens to repeated cycles of freezing and thawing in the laboratory by rapid freezing and thawing in water method.
STANDARD
· ASTM C666
FEATURES
◎ The compressor, heating device and power control part are all imported parts with high quality and high performance.
◎ When running at full load, the temperature difference at each point in the freeze-thaw cabinet does not exceed ±2℃.
◎ Display data curves in real time and automatically store test data.
◎ It has an automatic recovery function when the test is accidentally interrupted.
◎ Easy installation, small footprint, low noise
TECHNICAL PARAMETERS
TPDR-10
Condenser cooling system: Air cooling
Specimen size: 100×100×400mm. The size of the specimen for temperature measurement is the same, a temperature measuring hole is required in the middle
Capacity: 10 pcs
Freeze-thaw cycle: 2.5~4 hours, the thawing time is not less than 1/4 freeze-thaw cycle
Center temperature control: 8~-17±2℃
Liquid temperature control: High temperature: 18~23℃; Low temperature: -23~-18℃
Diagonal temperature control: The temperature difference in the cabinet does not exceed 1°C
Secondary protection temperature control: 50℃~-30℃
Temperature recorder: Touch screen
Freeze-thaw cycle medium: -45℃ antifreeze
Freeze-thaw cycle medium filling volume: The liquid level is 10-15mm higher than the specimen
Refrigerator compressor: Fully enclosed 3HP
Heating power: 3kW
Power supply: 3N380V 50Hz
Dimensions: 1750×800×1100mm
TPDR-16
Condenser cooling system: Air cooling
Specimen size: 100×100×400mm. The size of the specimen for temperature measurement is the same, a temperature measuring hole is required in the middle
Capacity: 16 pcs
Freeze-thaw cycle: 2.5~4 hours, the thawing time is not less than 1/4 freeze-thaw cycle
Center temperature control: 5~-18±2℃
Liquid temperature control: High temperature: 18~23℃; Low temperature: -23~-18℃
Diagonal temperature control: The temperature difference in the cabinet does not exceed 2°C
Secondary protection temperature control: 50℃~-30℃
Temperature recorder: Touch screen
Freeze-thaw cycle medium: -45℃ antifreeze
Freeze-thaw cycle medium filling volume: The liquid level is 10-15mm higher than the specimen
Refrigerator compressor: Fully enclosed 5HP
Heating power: 4.5kW
Power supply: 3N380V 50Hz
Dimensions: 1760×900×1100mm
TPDR-28
Condenser cooling system: Air cooling
Specimen size: 100×100×400mm. The size of the specimen for temperature measurement is the same, a temperature measuring hole is required in the middle
Capacity: 28 pcs
Freeze-thaw cycle: 2.5~4 hours, the thawing time is not less than 1/4 freeze-thaw cycle
Center temperature control: 5~-18±2℃
Liquid temperature control: High temperature: 18~23℃; Low temperature: -23~-18℃
Diagonal temperature control: The temperature difference in the cabinet does not exceed 2°C
Secondary protection temperature control: 50℃~-30℃
Temperature recorder: Touch screen
Freeze-thaw cycle medium: -45℃ antifreeze
Freeze-thaw cycle medium filling volume: The liquid level is 10-15mm higher than the specimen
Refrigerator compressor: Fully enclosed 6HP
Heating power: 6kW
Power supply: 3N380V 50Hz
Dimensions: Chamber 1450*1100*1700mm; Air cooling system 1950*900*1250mm